Immunity of cardiac pacemakers to
low frequency magnetic fieldS
A. Hedjiedj, M. Nadi
Laboratoire d’Instrumentation
Electronique de Nancy, Faculté des Sciences
Université H. Poincaré
de Nancy I – BP 239 – 54506 Vandoeuvre les Nancy – France
Abstract: This paper describes
the behavior of single and double chamber pacemakers subjected to a low frequency
magnetic field. The results allow to define the detection levels
of the stimulators and to show the reactions of the latter while the interfering
signal is applied.
INTRODUCTION
The important development of systems containing
electromagnetic sources, in everyday life or in industry (e.g. electrical appliances
or induction heating), has been accompanied by projected standards or regulations
aiming at avoiding the risks to patients with medical implants. In the case
of cardiac stimulators, studies dealing with interferences between low frequency
electromagnetic fields and these implants have been published. These studies
originally concerned interferences caused by the power distribution network
[1,2]. Since then, studies about the interferences between electronic article
surveillance systems and stimulators have also been published [3]. The main
difficulty of these studies is essentially linked to the great diversity of
possible situations, which makes comparing results difficult.
This article describes an in vitro study of the
behavior of cardiac stimulators submitted to low frequency electromagnetic disturbances.
The method developed is based upon the principle specific to electromagnetic
compatibility which recommends starting from the target (the cardiac stimulator),
identifying and quantifying the disturbances (the source), and then introducing
secondary influencing parameter in stepwise fashion. The tests are carried out
on single and double chamber stimulators, and the magnetic field is sinusoidal
at 50 Hz, 60 Hz, 10 kHz and 25 kHz. These tests allow to define the detection
levels of the stimulators and to reveal the reactions of the latter while the
interfering signal is being applied.
MATERIALS AND METHODS
The source of the disturbances is a Helmholtz-type
coil. It consists of two 1.5 m loops separated by a 0.75 m gap. The loops are
formed of six jointed turns of 16 mm2 wire. This structure, powered by a programmable
power generator, is used to subject a cardiac stimulator to a perfectly controlled
magnetic field, both in amplitude and direction, in order to simulate any real-world
exposure situation.
The stimulators tested are configured in inhibited
mode and both the stimulation and the detection are programmed in unipolar mode.
The device under test is made of a stimulator equipped with one or two probes
which, via the gel medium, forms one or two surface loops S (Fig. 1). Each probe
is completed by a rectangular loop made of 30 jointed turns, whose surface S’
is 165 cm2 (11 cm*15 cm). Using this loop allows to increase the apparent surface
of the loop formed by the stimulator and the probe. These loops are connected
in series with each probe (atrial and ventricular); the combination is placed
in a tank containing a gel that models the conductivity of tissues (for these
tests, the conductivity is 1mS/cm). Figure 2 shows the device under test in
the case of a single-chamber stimulator. The resistance R represents the load
on the stimulator.
Figure 1. The device under test
Figure 2. Increase of the apparent surface loop
The tank containing the stimulator is placed at the
center of the Helmholtz coil. The coupling between the magnetic field and the
loop formed by the device under test is maximal. The stimulator is submitted
to the magnetic field for 10 minutes. After each test, the use of telemetry
allows to measure the behavior of the stimulator during the application of the
interfering signal.
RESULTS
The results shown below illustrate how the two types
of stimulator work. Figure 3 represents a single chamber stimulator submitted
to a 25 kHz magnetic field; the detection sensitivity (S) is programmed to be
0.7 mV. The x-axis represents the VI voltage proportional to the
current flowing through the source. The vertical axes have a double reading
direction corresponding to the percentage stimulation (Stim) and detection (Det),
respectively. The tests highlight a window effect of the stimulator detection
circuits, within which numerous detections are noticed. There is an 80% maximum
detection ratio for VI=24.7 mV.
Figure 3. Behavior of a single chamber pacemaker
Figures 4 and 5 show the reactions of a double chamber
stimulator to 50 Hz and 25 kHz magnetic fields, respectively. The detection
sensitivity of the atrial and ventricular chambers are, respectively, SA=0.4
mV and SV=2.2 mV. The vertical axis on the left represents the percentage
of ventricular (VD) and atrial (AD) detections, and the periods in asynchronous
(As) and safety (Sec) modes. The axis on the right represents the percentage
of ventricular and atrial stimulations (V/A stim.). A window effect of the detection
circuits is again noticed. The reactions of the stimulator vary over the detection
range. Atrial and ventricular detections are noticed together with periods in
asynchronous and safety modes.
Figure 4. Behavior of a double chamber pacemaker
Figure 5. Behavior of
a double chamber pacemaker
Discussion
The results reveal the global behavior of the stimulators
during the application of the magnetic field. For the single chamber pacemaker
a detection corresponds to the inhibition of a pulse, which is translated
into a variation of the instantaneous frequency of the stimulator. For the double
chamber stimulator the problem is more complex and a thorough analysis is needed
to determine if the system works correctly.
It is possible to determine the equivalent magnetic field
corresponding to the detection levels without the additional(s) loop(s). Tables
1 and 2 show the equivalent magnetic field corresponding to the detection ranges
for the two types of stimulators under study, for the four studied frequencies.
By comparison with the recommended values of the European directive relative
to the exposure of the public to electromagnetic fields [4], these detection
levels remain in all cases superior to the maximum values.
Table 1 : Detection ranges of a double chamber pacemaker
|
Detection ranges
|
Sensitivity
|
50 Hz
|
60 Hz
|
10 kHz
|
25 kHz
|
SV=2.2
mV SA=0.4 mV
|
153 mT – 174 mT
|
142 mT - 161 mT
|
108 mT - 121 mT
|
47 mT
- 56 mT
|
Table 2 : Detection ranges of a single chamber pacemaker
|
Detection ranges
|
Sensitivity
|
10 kHz
|
25 kHz
|
0.7 mV
|
41 mT-51 mT
|
34 mT-43 mT
|
2 mV
|
108 mT-125 mT
|
79 mT- 86 mT
|
3 mV
|
-
|
109 mT-124 mT
|
References
[1] A. Scholten and J. Silny, 2001,The
interference threshold of cardiac pacemakers in electric 50 Hz fields, Journal
of Medical Engineering & Technology, Volume 25, Number 1, pages 1-11.
[2] Toivonen L., Valjus J., Hongisto M.,
Metso R., 1991, The influence of elevated 50 Hz electric and magnetic fields
on implanted cardiac pacemakers : the role of lead configuration and programming
of the sensitivity, Pace, Vol. 14, N°12, pp. 2114-2122.
[3] Lucas E.H., Johnson D., McElroy B.P.,
1994, The effects of electronic article surveillance systems on permanent cardiac
pacemakers : an in vitro study (Part II), PACE, Vol. 17, pp. 2021-2026.
[4] Journal officiel des communautés
européennes (7/07/1999) Recommandation du conseil du 12 juillet 1999 relative
à la limitation de l’exposition du public aux champs électromagnétiques (de
0 Hz à 300 GHz).
Acknowledgements: This work is
supported by EDF R&D.