![]() |
||||||||
![]() |
![]() |
2000, Volume 2, Number 2 | ||||||
|
Home
Current Issue
Table of Contents
Advanced Navigator Techniques
Abstract: The purpose of this study was to investigate and to optimize the performance of the real-time navigator technology on a clinical scanner for use in cardiac imaging. The studies involved experiments performed on phantoms and in vivo. The performance of the 2D RF pulses used for pencil beam excitation was found to be highly sufficient, provided a proper pulse design, accurately trimmed gradients and reasonable pulse parameters were used. The capability of the navigators for free-breathing, real-time gated coronary artery imaging was studied for several volunteers. Using a diaphragmatic gating window of 5 mm a reproducible image quality was achieved. To study spatial and temporal correlations of respiratory motion multiple navigator pulses were applied in different anatomical regions. For the correlation diaphragm-heart strong deviations from the linear model reported in literature were found. The results indicate, that a more complex model, including patient dependent hysteretic effects, might further improve the performance of real-time gating and prospective motion correction. Keywords: Navigator, coronary artery imaging, real-time gating, 2D RF pulses
IntroductionRespiratory motion can severely deteriorate the image quality of long cardiac-triggered MR imaging sequences. Therefore, gating based on navigator echoes [1,2] was introduced to reduce these artefacts. By means of a navigator, the position of the diaphragm can be monitored [3] and used as an input for the accept/reject decision of the gating algorithm. Furthermore, the navigator information may be used to perform prospective motion correction like slice-tracking to improve image quality and/or to allow a larger gating window [4]. On the other hand, robust real-time gating makes several demands on the performance of the navigator technique. The displacement of a selected anatomical region has to be monitored with high accuracy and destructive interference of the navigator pulse with the imaging volume has to be avoided. The navigator sequence should be short enough to allow versatile integration into MR imaging protocols. In addition, a fast evaluation of the measured displacements is necessary for real-time gating. Furthermore, an appropriate model for the respiratory motion of the heart is required, if prospective motion correction is performed. The purpose of this study was to investigate and to optimize the performance of a navigator on a clinical scanner with respect to these demands.MethodsIn vivo experiments with several healthy volunteers and phantom experiments were performed on a 1.5 T whole body scanner (GYROSCAN ACS NT, Philips Medical Systems) with self-shielded gradients (23 mT in 0.2 ms).Navigator pulse design
![]() Figure 1. Coronal survey with pencil beam through diaphragm (left). The Fourier transform of the gradient echo yields the projection of the pencil beam's magnetization onto the z-axis, the so-called navigator profile, which is sketched on the right for two different diaphragm positions. The contrast change between liver and lung results in a step-like shape. The displacement of the step with respect to a reference profile is determined in real-time by a cross-correlation algorithm. The total duration of the navigator pulse sequence including exciation, acquisition and evaluation is currently 20 ms.
Figure 2. Spiral trajectory of 2D-RF pulse (left) and corresponding point spread function (right). The aliasing rings appear due to the discrete k-space covering of the spiral. The central peak is used as a pencil beam navigator. Real-time gating
ResultsNavigator pulse performanceThe navigator performance was found to be highly sufficient for most applications. Using a simple eddy-current precompensation scheme for the gradient waveforms to correct for residual short-term eddy currents, an accuracy of positioning of the pencil beam better than 7 mm was achieved, which is only a fraction of a typical pencil beam diameter of 25 mm. Using strong gradient slew-rates (100 mT/m/ms) it was always possible to shift the aliasing rings out of the body without inducing off-resonance problems due to the longer pulse length (fig. 3). The spatial selectivity of the 2D-RF pulse was improved by proper weighting of the RF waveform. The navigator profiles obtained from the navigator echo show a well defined change in contrast at the lung-tissue interface (fig. 4). Hence, the displacement of the navigator profiles could be determined reliably by the cross-correlation algorithm. By sub-pixel interpolation an accuracy better than 1mm was achieved.
Figure 3. Transversal spin-echo images of the pencil beam excitation in the abdominal region (b-d). In the transversal survey (a) the expected positions of the central beam and the aliasing rings for a 3-turn spiral are indicated by white circles. With increasing number of k-space turns (b-d: 3,6,12) the rings are shifted out of the body. The corresponding pulse lengths are 1.7ms, 2.8 ms, and 5.7 ms, respectively.
Figure 4. Coronal survey indicating navigator positions (left), and corresponding navigator profiles (right). The red points indicate the respiration curves determined by cross-correlation in real-time. For the navigator through the heart a contribution from cardiac motion is superimposed. Coronary artery imaging
Figure 5. RCA (left) and LAD (right) of a healthy volunteer. The in-plane resolution of the 3D data set was 0.7 mm, the through-plane resolution 1.5 mm. Respiratory motion of the heart
![]() Figure 6. 2D histogram plots of the correlation heart-diaphragm for four selected volunteers. The corresponding navigator positions are shown in fig.4. Black bins indicate many counts, light bins indicate few counts. The red lines are linear fits to the data. ConclusionsPencil beam navigator pulses represent a powerful approach to monitor in vivo motion. When used for real-time gating high resolution coronary artery images can be obtained with reproducible quality during free breathing. The multiple navigator results indicate, that a proper model for the correlation diaphragm-heart, including patient dependent hysteretic effects, might further improve image quality or decrease scan time, when used for prospective motion correction. Multiple navigator pulses offer the potential for an automatic, patient dependent calibration of such a model.References[1] Liu, Y.L., Riederer, S.J., Rossmann, P.J., Grimm, R.C., Debbins, J.P. and Ehman, R.L., "A monitoring, feedback, and triggering system for reproducible breath-hold MR imaging", Magnetic Resonance in Medicine, vol. 30, p. 507, 1993.[2] Sachs, T.S., Meyer, C.H., Hu, B.S., Kohli, J., Nishimura, D. and Macovski, A., "Real-time motion detection in spiral MRI using navigators", Magnetic Resonance in Medicine, vol. 32, p. 639, 1994. [3] Wang, Y., Grimm, R.C., Felmlee, J.P., Riederer, S.J. and Ehman, R.L., "Algorithms for extracting motion information from navigator echoes", Magnetic Resonance in Medicine, vol. 36, p. 117, 1996. [4] Ehman, R.L. and Felmlee, J.P., "Adaptive technique for high-definition MR imaging of moving structures", Radiology, vol. 173, p. 255, 1989. [5] Pauly, J., Nishimura, D. and Macovski, A., "A k-space analysis of small-tip-angle excitations", Magnetic Resonance in Medicine, vol. 81, p. 43, 1989. [6] Nehrke, K., Börnert, P., Groen, J., Smink, J. and Böck, J.C, "On the performance and accuracy of 2D navigator pulses", Magnetic Resonance Imaging, vol. 17, no. 8, p. 1173, 1999. [7] Wang, Y., Riederer, S.J.
and Ehman, R.L.,"Respiratory motion of the hart: kinematics and the implications
for spatial resolution in coronary imaging", Magnetic Resonance in Medicine,
vol. 33, p. 713, 1995.
|
||||||||