Computer Modeling and Lead Field Theory
in the Analysis and Development of
Impedance Cardiography
P.K. Kauppinen*, J.A.K. Hyttinen*, T. Kööbi**, S. Kaukinen***, and J. Malmivuo*
* Tampere University of Technology/Ragnar Granit Institute, Tampere, Finland
** Tampere University Hospital/Dept. of Clinical Physiology, Tampere, Finland
***Tampere University Hospital/Dept. of Anesthesia and Intensive Care, Tampere, Finland
Correspondence: pasi.kauppinen@tut.fi
Abstract. Conventional impedance
cardiography (ICG) methods estimate parameters related to the function
of the heart from a single waveform, that reflects an integrated combination
of complex sources. Several modified ICG measurement configurations have
been suggested where for convenience or improved performance the standard
band electrodes are replaced with electrocardiography (ECG) electrodes.
However, measured data remain controversial, leading sometimes to errors
and discouraging utilization of ICG. Development of ICG has been mainly
empirical and its theoretical basis in terms of measurement sensitivity
is incompletely understood. We have developed computerised methods and
tools for calculating measurement sensitivity distributions of ICG electrode
configurations. In this report, the methods were applied to investigate
a) the sensitivity of the conventional and three modified ICG methods in
detecting regional conductivity changes and b) the prospects of developing
multiple aimed ICG recording configurations utilising the 12-lead ECG electrode
locations. The methods were applied with three anatomically realistic volume
conductor models: one based on Visible Human Man cryosection data and two
on magnetic resonance (MR) images representing end diastolic and end systolic
phases of the cardiac cycle. Preliminary clinical experimentation showed
logical correspondence with the simulations. The first results indicate
the applicability of the modelling approach in developing ICG measurement
configurations. However, the level of clinical relevance and potential
of the 12-lead method remains to be explored in studies employing more
dynamic modelling and acquisition several simultaneous ICG channels along
with invasive reference data.
Keywords: Impedance, Cardiography, Lead Field Theory, Sensitivity Distribution,
Computer Modeling, Visible Human Project, Electrode Configurations
Introduction
An ideal method of assessing information on
the cardiovascular system should be noninvasive, simple, atraumatic, inexpensive,
reliable and also applicable in long-term surveillance outside the cardiac
monitoring laboratory. Conventional ICG techniques provide a single impedance
tracing, from which parameters related to the pump function of the heart
are estimated [1-4]. Most of the properties of
ICG render it superior to other methods, the prominent exception being
its limited reliability, which has hampered its acceptance as a clinical
method. Most research on ICG has focused on comparison studies with reference
CO methods, producing widely scattered results with impedance-derived CO.
It has nonetheless been shown that in certain settings ICG can provide
useful information on the circulatory system. What is lacking for its more
widespread acceptance is thus that it be capable of producing quantitative
and reliable measurement.
The original ICG developed in the 1960s by Patterson et al.[1]
and Kubicek et al. [2] was based on a simple electrical
analogy from the thorax, with the assumption that the recorded impedance
variations measured on the surface of the thorax are essentially derived
from blood flow-related changes in the pulmonary vascular bed and arteries
resulting from the pumping action of the heart [5].
Although much work has been done to date, the development of ICG is still
principally based on this same ideology introduced decades ago, relying
strongly on an empirical approach. Several investigators have made modifications
to the CO equation or the electrode montage, yet no method has as yet yielded
reliable results. Continuing from the original development of ICG, use
of over-simplified models has limited and confined improvement in the technique
because of the gap between the model and the anatomy and physiology of
the system investigated. The refinements frequently presented in the literature
would not seem to constitute significant steps towards valid and stable
ICG method.
Time varying changes in impedance (ΔZ) reflect
the integrated combination of multiple sources, including tissue volume
and movement, tissue resistivity, blood distribution and blood flow changes
[6-8].
The contributions of these phenomena are reflected in ΔZ
depending on the electrode configuration used for the measurement [6,7,9-11].
Computer models designed to calculate the current flow in the thorax have
been used more recently in examining the ICG measurement configurations,
producing supporting data for the anticipated conception of the complexity
of the signal origin. Nevertheless no theoretical definition of ICG measurement
sensitivity distribution has thus far been provided. With numerical computer
models the anatomy can be accurately appraised, which should not only make
for a better understanding of the properties of existing ICG methods, but
facilitate the development of new ICG methods based more on theoretical
grounds.
A theoretical foundation exists for analysis of the measurement sensitivity
distribution of ICG based on the lead field theory introduced in 1953 for
bio-electric (i.e. ECG) analysis [12-14] and later
in the 1970s [15] for bio-impedance (BI) measurements.
BI measurement sensitivity distribution reflects how conductivity changes
throughout the volume affect the measured BI data. The application of lead
field theory in ICG has not evolved to its potential; only initial studies
with analytical models have hitherto been conducted to estimate the measurement
sensitivity distribution in cylindrically shaped objects with uniform conductivity
[16].
In this study, the lead field theory was utilized with computerized volume
conductor models of the human thorax in analyzing conventional ICG configurations
and in developing new ICG measurement configurations possessing more regional
measurement, i.e. sensitivity, properties.
Methods
Impedance Cardiography Sensitivity Distribution
Lead Vector, Lead Field and Reciprocity. The lead vector concept
explains the relationship between the electromotive source and the measured
lead voltage. Suppose that a single dipole source vector of magnitude
is at a fixed location in a linear, resistive volume conductor of arbitrary
shape and inhomogeneous conductivity. The measured unipolar lead potential, ,
corresponding to ,
is affected by the proportionality coefficient vector, having three orthogonal
components. This transfer vector is dependent on the location of the measurement
lead, the source dipole location, and the shape and the conductivity distribution
of the volume conductor. Now, by reason of the linearity assumption, applying
the superposition, the lead voltage can be expressed as the scalar product
of the transfer vectors between the measurement locations forming the lead
(e.g. locations a and b) and the dipole source as
|
 |
(1) |
where
is the 3-D transfer coefficient describing the sensitivity of the bioelectric
measurement at a specific location, the lead vector [17].
The lead field is a generalization of the lead vector; if the lead vectors
are mapped as a function of the source position throughout the volume conductor,
the single lead vectors will comprise a continuous vector field forming
the sensitivity distribution, called the lead-vector field or simply the
lead field.
If voltage measurement is made on the surface of an arbitrary volume
conductor, the measured signal in the lead arises from all the sources
in the conductor according to Eq. 1 at each location. The reciprocity theorem
states that in linear systems, applying a unit reciprocal current Ir
to the measurement lead gives rise to an electric field E in the
volume conductor, and the associated current density field
has exactly the same form as the lead field. If individual current dipoles
are characterized by current dipole moment per volume, ,
noting that ,
the expression for the lead potential becomes
|
 |
(2) |
where  denotes
the lead field [17,18].
As for the lead vectors, the lead field concept does not restrict the
complexity of the volume conductor; this can be of an arbitrary shape and
size with differing resistivities distributed in the volume.
Sensitivity Distribution of Bio-Impedance Measurement. The sensitivity
distribution of a BI measurement can be obtained by means of the lead field
theory. Now the source is not a current dipole within the volume conductor,
but the (varying) conductivity distribution. The sensitivity distribution
gives a relation between the impedance (and change in it) caused by a given
conductivity distribution (and its change). It describes how effectively
each region is contributing to the measured Z. If conductivity change
is not involved, the measured impedance Z is obtained with
|
 |
(3) |
where  and  ,
obtained with reciprocal energization, are the current density fields (i.e.
BI lead fields) associated with the current injection and voltage measurement
leads [17, 19]. This equation gives the contributions
from each region to the total impedance, and the dot product of the two
fields expresses the sensitivity of the measurement to conductivity changes
throughout the volume conductor. Effects of conductivity changes on measured
impedance can be calculated by  ,
where the time instants t1 and t2 refer
to situations before and after a conductivity change, with the assumption
that the changes in the lead fields are negligible due to the small conductivity
change.
The impact of a certain conductivity variation in different regions
depends on the sensitivity field. As the scalar field may possess positive
and negative values depending on the orientation of the two lead fields,
the measured impedance may either increase, decrease or be entirely unaffected
in consequence of a conductivity change in a particular region.
Utilising Eq. 3 with finite difference method (FDM) computer modelling,
information as to the respective capacity of different ICG measurements
to detect conductivity and its changes in the thorax can be estimated [20].
In the FDM, the modelled volume is divided into a three-dimensional resistor
network which reflects the thorax both geometrically and as a conductor.
Methods to construct and solve accurate volume conductor computer models
based on the FDM have been previously developed and validated [21].
The relative magnitude of the sensitivity field in a tissue type (or
a group of tissues considered as one target volume) gives a measure of
how conductivity variation in that tissue will affect the detected ΔZ.
The overall sensitivity of a tissue type is obtained by integrating the
sensitivity values of the tissue over the volume it occupies. This sensitivity
value can then be compared with the absolute total sensitivity of the model
as given by
|
 |
(4) |
where ng is the number of FDM elements in the target
volume and nt the number of tissue elements of a certain
type. The denominator is the sum of the absolute partial contributions
from all tissues (or tissue groups), and the numerator is the contribution
of the target tissue.
Volume Conductor Models
To yield useful information, the FDM model must employ the anatomy
of actual human structures. Three different anatomy models were employed
in the study:
Visible Human Man model. A particularly accurate source of anatomical
data, the U.S. National Library of Medicine?s Visible Human Man (VHM) [22-24]
was employed as basis for detailed FDM modeling. The original cryosection
images are 2048 by 1216 pixels in 24-bit colour, resulting in about 14
gigabytes of data in size. A total of 118 cryosection images from the top
of the head to the pelvis were segmented using a modified IARD (image enhancement,
amplitude segmentation, region growing, decision tree) volume segmentation
method which directly provides volume elements of anatomy data for FDM
mesh generation. Optional low-pass filtering, multiple amplitude segmentations,
region growing and decision trees were applied for the semiautomatic segmentation
procedure. Additional manual editing allowed classification of the smallest
details. This resulted in approximately 4 000 000 voxels and 32 segmented
and classified tissue types and organs. For data storage and image analysis
the accuracy of the images was reduced to 250x250 pixels using an 8-bit
gray scale colourmap. The resolution was from 0.044 to 5.7 cm3 for
the models utilised in the simulations.
Dynamic Model - Diastolic & Systolic models. The ECG triggered
end-systolic and end-diastolic MR image data sets used by Wang and Patterson
[8]
were segmented. A two-phase thorax model (i.e. two models of the same person
at different moments of the cardiac cycle) was constructed from these data
producing end diastole model (EDM) and end systole model (ESM). The number
of voxels was equal in both models, as the segmented outermost layer from
the first data set was used as base for segmentation of the other set.
Both models consisted of 70 slices and 30 tissue types. The resolution
of FDM elements in the ECG-triggered models varied from 0.10 to 5.8 cm3
resulting to 121431 elements.
Analysis of Conventional ICG
Contributions to the sensitivity distribution were assessed with the
VHM thorax model for four ICG electrode configurations utilizing conventional
band electrodes or modifications replacing the bands with spot electrodes:
a) Original configuration by Kubicek et al. using four band
electrodes [2]
b) Configuration by Penney et al. using four spot electrodes [25]
c) Configuration by Bernstein using eight spot electrodes [3]
d) Configuration proposed by Woltjer et al. using nine spot electrodes
[26].
Simulations were conducted to obtain the basal impedance, Z 0,
lead fields in the thorax generated by the current and the measurement
leads  and  ,
and the resulting measurement sensitivity distribution S. Electrode
configurations are shown in Fig. 1.

Figure 1. Placement of the electrodes on the VHM model for conventional
band and three alternative spot ICG electrode configurations. Locations
of the black spots are used for current injection and white spots for voltage
detection in actual measurements.
Development of 12-Lead ICG
The prospects of recording multiple ICG waveforms with more selective
sensitivity to particular regions of the thorax were investigated employing
the 12-lead electrode system. The lead field concept can be applied even
when several leads are combined. This facilitates synthesis of leads with
desired properties, for example, more specific leads to detect sources
induced by a heart disease.
The nine electrode locations of the 12-lead ECG electrode system were
used separately to calculate a basic set of lead fields for each model,
the VHM and two-phase models. A computer algorithm was developed to make
possible combinations with the 12-lead electrode system using at maximum
four electrodes at a time for either lead field in Eq. 3. Configurations,
which utilise the same electrode location for current injection and voltage
measurement, were omitted to reduce the skin-electrode impedance effect
on ΔZ. Deriving ICG measurement combinations
with the pre-calculated lead fields is a simple non-iterative calculation,
since the system is assumed to be linear. E.g. a lead field between the
chest leads V1 and V6 may be obtained by subtracting V1LL from
V6LL. On the other hand, the same result is obtained by subtracting
V1LA from V6LA. A total of 65476 impedance measurement
configurations utilising the 12-lead electrode locations was thus derived.
A database was computed for each model and 65476 measurement configurations
containing the information on the formation of Z0 and proportional
contributions according to Eq. 4. This was done for each classified tissue
listed and for a number of different tissue groups reflecting functional
structures of the cardiovascular system. Tissue groups were formed e.g.
from the tissues forming the systemic and pulmonary circulation in addition
to groups containing smaller number of tissues such as left atria together
with left ventricle. Further, the same calculations were applied to the
data produced by subtracting the sensitivity and Z0 values simulated
by the ECG-triggered models EDM and ESM.
Clinical Experiments
For the selected 12-lead-ICG measurements, first clinical experience
was acquired by making recordings in volunteers and valve patients in supine
position breathing spontaneously. The study involved twelve healthy volunteers
(age 30.5 ± 6.4 y mean ± SD, range 20 - 42 y; 11 male; 3
female, weight 79 ± 16 kg, 55 - 100 kg; height 179 ± 8.0
cm, 164 - 192 cm, BMI 24 ± 3.6 kg/m2, 17 - 30 kg/m2). The measurements
were also taken preoperatively on a group of 9 patients with valvular heart
disease (3 mitral, 6 aortic: age 58.4 ± 9.6, 35 - 72 y; all male;
weight 76 ± 14 kg, 62 - 111 kg; height 171 ? 2.8 cm, 166 - 175 cm,
BMI 26 ± 5.2 kg/m2, 20 - 39 kg/m2). Based on computer simulations
with the VHM and two-phase models, 237 measurement configurations derived
from the 12-lead system were selected as measurements to compare with the
simulations.
The measurements were performed by CircMonä
B202 (JR medical Ltd, Tallinn, Estonia), which includes an impedance channel
delivering 0.7 mA at 30 kHz. A novel software-controlled switching device
capable of electrically connecting the impedance measurement terminals
to any number of applied electrodes was used in combination with the CircMon
[27].
The electrode configuration used for impedance measurement could thus be
altered rapidly by computer control without manual operation. A period
of 10 s was measured with each configuration.
Results
Conventional ICGs
The conventional band electrode configuration and alternative configurations
suggested to replace the bands were shown not to be specifically sensitive
in measuring conductivity changes in regions generally considered important
in measuring CO, namely the heart, lungs or aorta and other large vascular
trees. More than half of the measurement sensitivity in each case studied
was concentrated in the skeletal muscle. Furthermore, the results showed
heterogeneous current field flow in the thorax, and modifying the electrode
configuration resulted in different sensitivity distributions which must
have an influence on the composition of measured signals. Sensitivity distributions
are visualised in Figure 2.

Figure 2. Mid-frontal and transversal views of sensitivity field distributions.
a) conventional band, b) four-spot, c) eight-spot and d) nine-spot configurations.
Zero sensitivity is indicated with black colour, positive sensitivities
are visualized with hot colourmap and negative with cool colourmap. Average
sensitivity values at viewing planes were used to scale the colourmap range
for each case to obtain added brightness.
12-Lead ICG
Simulated electrode configurations indicated markedly different measurement
sensitivity distributions between each other. As compared to the conventional
ICG, clearly enhanced sensitivities were obtained in various tissues. However,
no absolutely selective measurement configurations for particular structures
of the cardiovascular system were obtained. As a consequence, no measurements
applying any combination of the 12-lead electrodes will produce exclusive
data from a particular region, although the partial contributions from
various regions may be significantly increased.
Clinical Experiments
Clinical experiments with the ICG measurements derived from the 12-lead
electrode system showed logical correspondence to the simulations, supporting
the theoretically predicted differences between the configurations. Systolic
ΔZmax
correlated significantly to the difference in basal impedances simulated
with the two-phase model, but logically, no significant correlation was
found to the VHM or either of the two-phase models when considered separately.
The strongest correlations between sensitivity and measured waveform were
noted for the right ventricle and the area under the systolic part of the
impedance curve. Recorded 12-lead signals had characteristic waveforms
and landmarks not coinciding with those of conventional ICG, indicating
varied information content between the configurations. Furthermore, configurations
were noted showing a suggestive resemblance to invasive data and morphological
variations in disease. Fig. 3 shows example averaged waveforms recorded
from both study populations.

Figure 3. Examples of the 12-lead based ICG recordings shown as average
signals from the study groups (volunteers and valvular patients). a) Tracings
with small inter-group variation, b) changes between the groups in the
time instant of the maximum impedance deflection, c) characteristic signals
with notable peaks or deflections missing between the groups, d) large
inter-group MAPEs. Tracings marked i and ii together identify the study
populations.
Conclusions
Using ICG for measuring CO with conventional methods involves many
assumptions and simplifications. The technique is by nature particularly
indirect, reflecting complex simultaneous variations in the electrical,
geometrical and physiological properties of tissues. To derive valid measures
of cardiac-related parameters using BI measurements it is of primary importance
that the signal contain the relevant information in a distinguishable manner.
In this series of studies, the conventional ICG and a large number of measurement
configurations derived from the 12-lead ECG electrode system were analyzed
in respect of their measurement sensitivity distributions. The uniqueness
of the methods used was that the sensitivity of the whole measurement setting
was obtained in accurate models with a single simulation without approximations
of quantitative conductivity changes occurring in the thorax. This was
important in that an understanding of the ICG requires a conception of
how the sensitivity of measurement is distributed, not only how large is
the global impedance change due to a particular resistivity change in a
specific organ.
Conventional ICGs
Simulation results emphasized the multiregional sampling sensitivity
of the studied ICG configurations. For the conventional ICGs, only an approximately
5 % contribution from all blood masses and cardiac tissue was detected.
This can be taken to imply that the valuable information, i.e. the information
needed to determine the CO or other desired parameters, overlaps with a
wide range of other information unretrievable from ΔZ.
Although useful information has been obtained from the ICG waveform, it
originates from a multiplicity of sources with unpredictably varying contributions
depending on the characteristics and hemodynamic condition of the subject.
Thus, particular caution is called for when applying ICG to clinical work.
It is unlikely that a universally ideal electrode configuration providing
accurate measurements exists for ICG. If the number of unpredictable factors
contributing to or modifying the ICG waveform are many, at least as many
specific measurements should be taken as there are contributing factors
to establish the state of the system.
12-Lead ICG
Numerical modeling with the lead field theoretical approach made possible
detailed analysis of a large number of configurations, in this study 65
476 derived from the 12-lead ECG electrode system. Increasing the contribution
from a limited region may improve the physiological relevance of recorded
data, which was achieved in theory with the regional multi-electrode measuring
configurations. According to simulations with the VHM and the two-phase
models, greatly increased sensitivities were obtained for each classified
tissue with certain measurement configurations. For the tissues of the
cardiovascular structures, a maximum of 75 % proportional sensitivity was
attained. For the aortas and vena cavas the values were relatively small,
since the electrode locations of the 12-lead system are not favorable for
vertical measurements, especially when the right leg, often used in ECG
acquisition, was ignored in the simulations. For this reason, the maximal
sensitivity for the pulmonary circulation was twice that of the systemic
circulation. Although highly elevated sensitivities were obtained, it was
not possible to achieve fully selective measurements for any of the tissues.
Moreover, even with these enhanced measuring configurations, everything
still affects everything; nonetheless their relative contributions should
be more favorable than in conventional methods to produce regional information.
Computer Models
The anatomical differences between the VHM and the two-phase models
were considerable; for instance, the total volumes were 47 and 21 l for
the VHM and the two-phase models, respectively. In spite of this, many
of the most selective configurations for certain anatomical region or tissue
were identical independent of the model applied. This result was unexpected,
since to achieve high sensitivity in a certain region and low in others
requires in principle a measuring configuration where lead fields of current
and voltage electrodes are practically perpendicular to each other. Slight
deviations in model geometry or electrode locations could then markedly
modify the partial sensitivity values. On the other hand, the derived configurations
producing high sensitivities utilized more than four electrodes. Nonetheless
these configurations with high sensitivity produced small basal impedances,
which supports the conception that perpendicular electrode placement produces
zero basal impedance. Also, as each tissue was considered as a whole in
the analysis, the contribution from large tissues such as skeletal muscle
could have been reduced by the fact that both positive and negative sensitivity
values present can result in practically nil total sensitivity in that
tissue. This, of course, does not imply that the tissue has no effect if
its conductivity changes. On the other hand, however, the contribution
from skeletal muscle has been noted to be almost nil when measured with
conventional band electrodes placed on the abdomen [7, 28].
Clinical Experiments
Recorded 12-lead signals had characteristic landmarks not coincing
with those of conventional ICG, indicating varied information content between
the configurations. Furthermore, signals were noted showing a suggestive
resemblance to invasive data and morphological variations in disease not
present with conventional ICG. Valvular disease was detected when investigating
at least two signals simultaneously, a single ICG signal cannot produce
information for the identification of the existence of the disease. An
important limitation in the clinical measurements is the instrumentation
restricting the analysis of collected signals since only one channel can
be recorded at a time. To collect a wide range of clinical data from patients
during invasive measurements, the recording system should be implemented
in a multi-channel form allowing parallel recording of several independent
12-lead ICG signals.
Conclusions
The findings demonstrated that the lead field approach is suitable
in ICG, allowing transition from the empirical selection of a few parameters
from empirically measured impedance waveforms to taking full advantage
of the information embedded in a multiplicity of regional BI measurements.
On the theoretical side, extending the two-phase to a multi-phase model
one could simulate the ΔZ waveform and investigate
geometrical and conductivity changes separately and not only assess the
sensitivity distribution at one time instant. Comparisons with clinical
data would facilitate the development of models and provide a guide in
selecting valid tissue conductivities.
Ideally, it would be desirable to achieve measurement sensitivity only
in the region of interest, producing selective information. Such measurement
would require lead fields having null values anywhere else than in the
target region, which is practically impossible for any surface electrode
system. However, employing multiple measurements with different and known
sensitivities to the relevant component in the system may convey useful
information related to some specific event or region undetectable by conventional
ICG methods. Presently, however, the lack of appropriate instrumentation
restricts this approach.
References
[1] Patterson, R. P., Kubicek, W. G., Kinnen, E., Witsoe,
D. A. and Noren, G., "Development of an electrical impedance plethysmography
system to monitor cardiac output", presented at 1st Ann Rocky Mountain
Bioengineering Symposium, 1964.
[2] Kubicek, W. G., Karnegis, J. N., Patterson, R.
P., Witsoe, D. A. and Mattson, R. H., "Development and evaluation of an
impedance cardiac output system", Aerosp Med, vol. 37, pp. 1208-1212, 1966.
[3] Bernstein, D. P., "A new stroke volume equation
for thoracic electrical bioimpedance: theory and rationale", Crit Care
Med, vol. 14, pp. 904-9, 1986.
[4] Kööbi, T., Kaukinen, S., Turjanmaa,
V. M. and Uusitalo, A. J., "Whole-body impedance cardiography in the measurement
of cardiac output", Crit Care Med, vol. 25, pp. 779-85, 1997.
[5] Kinnen, E., Kubicek, W. G., Hill, D. W. and Turton,
G., "Thoracic cage impedance measurements: impedance plethysmographic determination
of cardiac output (An interpretative study)", U.S. Air Force School of
Aerospace Medicine, Brooks Air Force Base, Texas, vol. SAM-TDR-64, pp.
12, 1964.
[6] Geddes, L. A. and Baker, L. E., "Principles of
applied biomedical instrumentation", 2nd ed. New York: John Wiley &
Sons, 1975.
[7] Mohapatra, S. N., "Non-invasive cardiovascular
monitoring by electrical impedance technique", London: Pitman, 1981.
[8] Wang, L. and Patterson, R., "Multiple sources
of the impedance cardiogram based on 3-D finite difference human thorax
models", IEEE Trans Biomed Eng, vol. 42, pp. 141-8, 1995.
[9] Patterson, R. P., Wang, L., Raza, B. and Wood,
K., "Mapping the cardiogenic impedance signal on the thoracic surface",
Med Biol Eng Comput, vol. 28, pp. 212-6, 1990.
[10] Kauppinen, P. K., Hyttinen, J. A. and Malmivuo,
J. A., "Sensitivity Distributions of Impedance Cardiography Using Band
and Spot Electrodes Analysed by a 3-D Computer Model", Ann Biomed Eng,
vol. 26, pp. 694-702, 1998.
[11] Kauppinen, P. K., Hyttinen, J., Kööbi,
T. and Malmivuo, J., "Lead field theoretical approach in bio-impedance
measurements - towards more controlled measurement sensitivity", Ann N
Y Acad Sci, vol. 873, pp. 135-142, 1999.
[12] McFee, R. and Johnston, F. D., "Electrocardiographic
leads I. Introduction", Circulation, vol. 8, pp. 554-568, 1953.
[13] McFee, R. and Johnston, F. D., "Electrocardiographic
leads II. Analysis", Circulation, vol. 9, pp. 255-266, 1954.
[14] McFee, R. and Johnston, F. D., "Electrocardiographic
leads III. Synthesis", Circulation, vol. 9, pp. 868-880, 1954.
[15] Geselowitz, D. B., "An application of electrocardiographic
lead theory to impedance plethysmography", IEEE Trans Biomed Eng, vol.
18, pp. 38-41, 1971.
[16] Penney, B. C., Narducci, L. M., Peura, R. A.,
Anderson, F. A. J. and Wheeler, H. B., "The impedance plethysmographic
sampling field in the human calf", IEEE Trans Biomed Eng, vol. 26, pp.
193-8, 1979.
[17] Malmivuo, J. and Plonsey, R., "Bioelectromagnetism:
principles and application of bioelectric and biomagnetic fields", New
York: Oxford University Press, 1995.
[18] Plonsey, R., "Bioelectric Phenomena", New York:
McGraw-Hill Book Co., 1969.
[19] Kauppinen, P., "Measurement properties of impedance
cardiography -- FDM computer model studies", Lic.Tech. Thesis. Tampere,
Finland: Tampere University of Technology, 1997.
[20] Kauppinen, P., "Application of lead field theory
in the analysis and development of impedance cardiography", PhD Thesis.
Tampere: Tampere University of Technology, 1999, pp. 152.
[21] Kauppinen, P., Hyttinen, J., Laarne, P. and
Malmivuo, J., "A software implementation for detailed volume conductor
modelling in electrophysiology using finite difference method", Comput
Meth Prog Bio, vol. 58, pp. 191-203, 1999.
[22] Ackerman, M. J., "The Visible Human Project",
J Biocommun, vol. 18, pp. 14, 1991.
[23] Lindberg, D. A. and Humphreys, B. L., "Computers
in medicine", JAMA, vol. 273, pp. 1667-1668, 1995.
[24] Spitzer, V., Ackerman, M. J., Scherzinger, A.
L. and Whitlock, D., "The visible human male: a technical report", J Am
Med Inform Assoc, vol. 3, pp. 118-130, 1996.
[25] Penney, B. C., Patwardhan, N. A. and Wheeler,
H. B., "Simplified electrode array for impedance cardiography", Med Biol
Eng Comput, vol. 23, pp. 1-7, 1985.
[26] Woltjer, H. H., Arntzen, B. W., Bogaard, H.
J. and de Vries, P. M., "Optimalisation of the spot electrode array in
impedance cardiography", Med Biol Eng Comput, vol. 34, pp. 84-7, 1996.
[27] Kauppinen, P. K., Hyttinen, J. A. K., Kööbi,
T. and Malmivuo, J., "Multiple lead recordings improve accuracy of bio-impedance
plethysmographic technique", Med Eng Phys, vol. 21, pp. 371-375, 1999.
[28] Nyboer, J. "Electrical impedance plethysmography",
2nd ed. Springfield: Charles C Thomas, 1970.
|