L. Guerri*, P. Colli Franzone**, M. Pennacchio*** and B. Taccardi****
* Department of Mathematics, University of Piemonte Orientale, Alessandria, Italy
** Department of Mathematics, University of Pavia, Pavia, Italy
*** Institute of Numerical Analysis - CNR, Pavia, Italy
**** CVRTI - University of Utah, Salt Lake City, UT, USA
Introduction
The excitation process in the myocardium, the associate potential distribution
and the electrograms are influenced by the fiber structure and the anisotropy
of the myocardial tissue. In fact propagation is faster along than across
fiber, and positive potentials are more frequently observed in those areas
toward which excitation spreads along fibers. As a consequence, the shape
of the unipolar electrograms (EGs) depends on whether the recording site
is reached by an excitation wave moving along or across or obliquely to
the fiber direction [13]. Only few simulations of
EGs based on the full anisotropic bidomain model have appeared in literature
[5,14,1,10,21]
A preliminary study of the effect of the reference potential on unipolar
electrograms (EGs) in a large slab of myocardium was carried out in [21].
In this paper we use a mathematical model of the left ventricle, based
on the anisotropic bidomain, taking into account the intramural rotation
and obliqueness of the fibers and a simplified Purkinje network. The computation
of EGs requires a refined numerical technique to avoid spurious oscillations
and peaks. This adaptive procedure, which has some affinity with that proposed
in [5], was first used in [2]
and was adapted to the model of this study. The analysis of the EG shape
is facilitated by the use of the splitting technique applied to sources,
potential fields and EGs decomposing them into axial and conormal components.
Also the effect of the drift of the reference potential must be considered.
Methods
Mathematical Model
The numerical simulations were carried out on a monoventricular model
previously used to simulate the spread of excitation and the associated
potential distributions. The model, representing the left ventricle as
an ellipsoid symmetric with respect to the z-axis and truncated at the
basis and the apex, takes into account the fiber rotation counterclockwise
(CCW) from epicardium (-45o) to endocardium (75o).
and the epi- endocardial obliqueness see [18]. Thus,
the fibers do not lie on the nested ellipsoidal surfaces but intersect
them at a small angle. For more details on the geometry and on the fiber
architecture of our model of the left ventricle see [3].
The ventricular model, used in this simulation study, incorporates the
main electrical and structural features of the myocardium such as :
-
a) the anisotropic electrical conductivities of the intra- and extra-cellular
media with unequal anisotropic ratio;
-
b) the curvature and variable thickness of the ventricular wall;
-
c) the intramural fiber rotation and obliqueness, whose combined effect
results in fiber pathways developing on nested toroidal surfaces inside
the ventricular wall (see [3,18]);
-
d) a simplified subendocardial Purkinje network;
-
e) two fluid layers, one having conductivity similar to the blood and lining
the endocardium, the other in contact with the epicardial surface, with
conductivity similar to the average conductivity of chest tissues (i.e., 2
).
In our model the ventricular cavity is not completely filled by blood and
the fluid layer in contact with the epicardium has a thickness varying
from 2.5 to 5 mm, not comparable with that of the conducting fluid filling
the experimental tank. Nevertheless, the model is a good approximation
of the experimental setup related to isolated dog hearts, especially for
epicardial and intramural observation points.
Therefore, most of the essential factors affecting the EGs are present
in both the experimental and the modeling framework. This allowed us to
perform a qualitative comparison between measured and computed EGs.
We briefly recall that the mathematical model used to compute EGs is
based on the bidomain approach where two superposed and connected media,
the intracellular (i) and the extracellular (e), are considered [7,9].
In the following we shall denote by H the heart tissue, by
Ωb and
σb
the cavitary blood and its conductivity, by
Ωf and
σf
the fluid volume adjacent to the epicardium and its conductivity, by
the whole extracardiac medium in contact with H and by
.
Let
,
,
i.e.
represents the parts of the epi and/or endocardium in contact with
,
and
,
,
hence
and
.
In particular the epicardial surface will be denoted by
.
We define M=Mi+Me the ``bulk''
conductivity tensor and
with M0 conductivity tensor in
given by
where I denote the identity matrix. Let
and
,
i.e.
represents the parts of the epi and/or endocardium in contact with
.
In particular the epicardial surface will be denoted by
.
The bidomain is characterized by:
-
a) the local fiber direction
and the global fiber structure
-
b) the conductivity coefficients
along and across fiber in the (i), (e) media.
is the same for all directions orthogonal to
.
The conductivity tensors
are defined by:
 |
(1) |
where

is a column vector and

its transpose. In the bidomain model the following potentials are considered:

,

and

representing
respectively the intra-, extracellular transmembrane and extracardiac potentials
and the current vector densities associated to the intra, extracellular
and extracardiac potentials :
Moreover setting
 |
(2) |
and applying the current conservation to the bidomain model we obtain
:
 |
(3) |
Assuming known the transmembrane potential
then in terms of the extracellular or extracardiac potential
the bidomain current conservation equations (3) are
equivalent to :
 |
(4) |
where
denotes the jump of
through
,
i.e.
with
the traces taken on the positive and negative side of
with respect to the oriented normal. Therefore problem (4)
provides the extracellular or extracardiac potential
from the knowledge of
,
see ([2],[4]). As a consequence
of the bidomain approach and assuming known the transmembrane potential
distribution v, the term
plays the role of the ``impressed'' or ``source'' current density generating
the extracellular potential. Different mathematical approaches can be used
to simulate EGs (see e.g. [6,8,15,21]).
Because EGs are usually recorded from a limited number of sites, an integral
formulation (see [23,7]) is
more convenient than differential or variational representations, for large
scale simulations as required for an accurate computation of extracardiac,
epicardial or intramural EGs [15,2].
Each EG represents the time course of the potential difference between
one point of the domain, called observation point, and a ``reference potential''.
For unipolar EGs, the reference potential is that at a remote site or,
more frequently, is obtained by averaging the potential values over a set
of three or more points or over a surface, e.g. the entire insulated surface
of the conducting volume as suggested in [17]. Recent
experimental observations with isolated canine hearts in a torso-shaped
tank showed that the potential of Wilson's central terminal is close to
the average potential [19] .
In this work EGs are computed by using the latter reference potential,
i.e. the average potential on the epicardial surface
.
Then we consider
 |
(5) |
For
,
we introduce the solution
of the following elliptic problem :
|
(6) |
with
characteristic function of
,
i.e.
and
denotes the jump of
through
,
i.e.
with
the traces taken on the positive and negative side of
with respect to the oriented normal.
Applying the second Green formula to the couple
in
and in H with
solution of (6), adding these two relations we obtain
the following integral representation:
 |
(7) |
The boundary condition imposed for the ``lead field'' actually reflects
the property that
,
defined by (5), has zero average on the epicardium.
The same integral representation can be extended for observation points
,
i.e. within the wall or in the extracardiac medium. In these cases the
lead field
is defined by:
|
(8) |
For more details on the integral representation of w see [2]
.
The integral approach easily allows to change the reference potential
of the EGs without recomputing them. For instance let us consider, as suggested
by Spach et al. in [17], as reference potential the
average potential on the whole insulated surface
.
If we denote by
 |
(9) |
with
solution of:
 |
|
|
(10) |
then it is easy to verify that
has zero average on the whole surface
.
Thus, from (7) and (9) we have
and, defining
,
we obtain that
is solution of
 |
|
|
(11) |
In conclusion, defining
,
we have
 |
(12) |
hence to change the reference potential fixed for the EGs, i.e. in this
case considering the average potential on the whole
instead of the average potential on
,
we have to compute
and
and
to add wd(t) to
.
Note that
,
solution of problem (11), can be computed only once
since it does not depend on the observation point
;
moreover
is
a smooth function without singularity, i.e. with no delta function as the
lead fields (6) and (10);
hence it can be computed using a uniform mesh without refining it near
the observation point
(see [2] for more details). This procedure can easily
be applied to obtain other reference potentials, e.g. the average on a
subset of
or
.
In a previous paper [2], we developed and validated
an efficient numerical technique for the computation of EGs, at any point
inside or outside the myocardium, free from spurious oscillations and peaks.
We now want to use this numerical procedure to elucidate the mechanisms
that produce the frequently observed multiphasic EG waveform; more generally,
we want to analyze the factors that influence the shape, amplitude and
polarity of unipolar EGs.
Source splitting
We briefly recall the main formulae related to the split form which
will be applied to the analysis of the EG wave shape. The ``bulk'' conductivity
tensor is defined by M=Mi+Me.
Setting
it
is easy to verify that:
 |
(13) |
and that the case of equal anisotropy ratio
is characterized by
.
Using (13), since
we obtain:
 |
(14) |
where
 |
(15) |
We remark that
,
are distributed current densities with dipole axes parallel respectively
to the fiber direction
and to the conormal vector
.
For this reason
and
are called
axial and conormal current densities. From (7)
it follows that :
 |
(16) |
where
 |
(17) |
 |
(18) |
For t fixed and
variable (16) defines a split of the potential distribution
into the axial and conormal component of the field, while
for
fixed and t variable (16) defines a split of the
EG into the axial and conormal EG waveforms.
We recall that for a bidomain with equal anisotropy ratio we have
and consequently in our decomposition (16) only the conormal
component is present, i.e. the full EG
reduces to the conormal EG component
.
Assuming
constant in H, i.e.
,
which holds if
are constant, using the Green formula
and taking into account the properties of
in (6) we obtain the following further decomposition
of the conormal field
:
 |
(19) |
where
 |
(20) |
where
denotes the epicardial surface area.
The component uJ is usually called ``jump component''
because its effect on the conormal potential wc reduces
to a shift proportional to the jump of the transmembrane potential from
the resting to the plateau value through the excitation front for t fixed
or at the activation time for
fixed.
We note that
is independent of the observation point
where the EGs are simulated, positive and increasing with time. Therefore,
it needs to be computed only once and can be considered as a drift component
related to the chosen reference potential; we call this term the drift
of the conormal component of the EG.
We remark that
contributes to the conormal potential
only when the heart is in contact with an extracardiac conducting medium
(i.e.
)
and depends on the trace on
of
the transmembrane potential
.
Since the component (20) represents the potential generated
by a conormal dipole layer lying on the non-insulated heart surface
we refer to this field component as usual (see e.g. [6,7,15])
as heart surface source model, i.e. this component is the potential
generated by a double layer on
with dipoles parallel to
and moment proportional to the transmembrane potential v.
In conclusion we get:
 |
(21) |
For a derivation of the split form under more general conditions and
for more details see [] and [2]. In the special
case of H fully insulated (i.e.
and
)
and
constant the conormal component reduces to:
 |
(22) |
Computation of
The integral formulation (7) requires the knowledge
of the time-space distribution of the transmembrane potential
.
Our numerical procedure can be used to compute EGs over the whole cardiac
cycle from ventricular excitation to recovery; however in this paper we
limit our simulations to waveforms associated with the depolarization phase,
i.e. to the so called QRS complex of EGs.
We use the eikonal model (see [1,3])
to build an approximation of the transmembrane potential
;
in this approximation, assuming uniform membrane excitability properties,
is given by:
 |
(23) |
where
-
ii)
-
the activation time
is obtained by solving the eikonal equation (see []) describing the motion
of the excitation wave front;
-
iii)
-
the action potential upstroke is given by:
 |
(24) |
with Cm membrane capacitance per unit area, vr,
vp,
vthresting,
plateau and threshold value of the transmembrane potential
v, Gmaximum
membrane conductance per unit area and s
= (vp-vr)/(2(vth-vr))-1
(see []) .
Numerical approximation
The numerical computation of the integral (7) is carried
out by FEM and requires special care because of the singularity of
at the observation point
and the presence of the moving steep wave front characterizing
during
the depolarization phase.
We sketch here the main steps of the procedure used to improve the accuracy
of
:
-
b)
-
refinement of the Finite Element Method mesh in computing the lead field
near its singular point
.
Since we know that
is a smooth function far from
we
build a grid having small elements in proximity of the singularity of
and elements of increasing dimensions further away. A structured grid refinement
around the singularity was performed maintaining approximately the same
number of elements of the mesh associated to the computation of the activation
time
so that the computer time and memory remain unchanged.
-
c)
-
at each time instant, identification of the elements crossed by the
wave front and successive subdivision of these elements;
-
1.
-
only the elements of the 3-D grid crossed by the wave front are processed
in the computation of the integral (7) on H.
To identify these elements we fix
(e.g.
)
and we determine
For the profile (24) of the action potential upstroke
it is easy to verify that
with
and
i.e.
vanishes to a good approximation. Thus the contribution of an element to
the integral is neglected if
for
all the grid nodes
of the element. Otherwise we assume that the element is crossed by the
wave front.
-
2.
-
the contribution of these elements to the integral is evaluated subdividing
them into
sub-elements
where
indicate the number of subdivisions performed on the element with respect
to the
directions for the ellipsoidal ventricle. The volume integral is computed
by successively adding the contributions of the grid-sub-elements crossed
by the wave front at a given time instant. In our simulations a typical
subdivision was defined by
,
and mr=1 and the numerical evaluation of the integral
has been carried out on each sub-element by 1-node Gaussian formula.
-
d)
-
nonlinear interpolation of
;
The activation time
,
a smooth function without sharp variations, is well approximated on each
sub-element using a trilinear polynomial
defined by the values of
at the nodes of the macro hexahedral element. Hence using this trilinear
interpolation
by means of (23) we are led to consider, for the function
exhibiting high gradient, the following nonlinear, but more accurate, interpolation
formula:

-
e)
-
use of the split form (16) of
;
The
axial
and conormal component wa, wc can be
computed by exploiting the procedure outlined in steps b), c) for the evaluation
of the volume integral (17) and (18)
respectively. Moreover, the accuracy of the computation of w can
be increased in the case of H fully insulated by combining the split
form (22) with the numerical technique outlined above.
In fact, from (22), the conormal component
can
be evaluated by computing the surface integral
and the jump
by using (23). For the evaluation of the surface integral
we used steps b), c) adapted to surface elements. In conclusion in the
fully insulated case we apply the sub-element technique to evaluate the
volume integral (17) yielding
and we compute
using (19).
In this way it is possible, by using the outlined numerical procedure,
to obtain EGs free from numerical artifacts like spurious oscillations
or peaks, with a limited computational effort, as established in [
2].
Results
Experimental EGs showed that within 10-15 mm from the pacing site the
shape of unipolar EGs markedly depends on whether excitation reaches the
recording site by traveling mainly along or across fibers [16,20,21].
In the following, these recording sites will be referred to as along-fiber
or cross-fiber sites. Fig. 1 shows the morphological differences
in epicardial EGs recorded from 4 sites located just beyond the 18 msec
isochrone, along longitudinal (along-fiber) or transverse (cross-
fiber) pathways. Due to higher longitudinal conduction velocity, the
18 msec isochrone is elongated, quasi-elliptical, centered about the pacing
site, with major axis oriented along the local direction of epicardial
fibers.

Figure 1. Isolated dog heart immersed in an electrolytic
tank shaped as a 10-year boy's torso, filled with a 500
cm solution. CX = circumflex coronary branch. LAD = left anterior descending
coronary branch. Ventricle was paced from star-marked site. Elliptical
contour is the 20 ms isochrone. Right side : unipolar EGs were recorded
from sites denoted by large dots near the epicardial elliptical isochrone
at 18 msec after pacing . Left side: computed EGs in monoventricular model
with the epi- and endocardium in contact with conducting media. Epicardial
EGs No 1 through No 4 were computed for sites marked by dots around the
simulated epicardial isochrone at 20 msec after stimulus.
We define the EGs as monophasic, biphasic etc. according to the number
of relative maxima and minima that occur before and immediately after the
intrinsic deflection (the major downstroke).
The EGs displayed in Fig. 1 show that:
-
the cross-fiber EGs are monophasic. The portion preceding the intrinsic
deflection is monotonically negative-going, as previously described in
[19,20,21];
-
the along-fiber EGs are biphasic, with an initial positive R-wave
followed by the negative-going intrinsic deflection [22].
At later stages of ventricular excitation, when approximately 50

of the ventricular surface has been excited, the experimental EGs, recorded
from sites close to the
60 ms isochrone, (Fig. 2 ), are biphasic,
with the positive and negative phases having almost equal magnitude.

Figure 2. Measured and computed epicardial electrograms.
Measured and simulated epicardial isochrones at 60 and 80 msec after the
stimulus are traced in the right and left Panel respectively. Same layout
and explanations as in Figure 1.
For a positive and negative R and S wave respectively
we introduce the ratio between the R wave amplitude and the pick
to pick drop given by:
.
The progressive positive trend of the
wave
amplitude ratio, which reaches its maximum value in Fig.2, is an effect
of the drift of the reference potential, which moves from near the positive
to near the negative extreme of the potential range during ventricular
excitation, as recently recognized and discussed in our previous work [19,21].
These studies showed that all unipolar EGs recorded directly from the heart,
when referenced to Wilson's Central Terminal or to the average epicardial
potential, contain a positive-trending component, associated with the drift
of the reference potential. This component causes the
ratio
to increase progressively from zero to 1 as a function of the amount of
excited epicardial area.
As a result, EGs recorded from areas that are excited at the end of
the QRS interval, near the extinction region, are entirely positive (Fig.
3 ).

Figure 3. Measured and computed epicardial electrograms.
Measured and simulated epicardial isochrones at 85 and 110 msec after the
stimulus are traced in the right and left Panel respectively. Same layout
and explanations as in Figure 1.
Proceeding from the pacing site along cross-fiber pathways, the
EGs shift from monophasic to tetraphasic. These multiphasic waveforms occur
mostly at the epicardial and subepicardial levels in experimentally recorded
EGs, in regions where excitation spreads across fibers. In regions located
outside the 40 msec isochrone (see Fig. 4):
-
the along-fiber EGs still have a biphasic shape with an R wave preceding
the intrinsic deflection.
-
the cross-fiber EGs show a more complex time course; in fact we
have:
-
EGs that, after an initially negative- going phase, show a minimum followed
by a maximum before the intrinsic deflection (Fig. 4 A, EG No 1).
-
EGs that, after a flat initial phase, show a positive- negative wave followed
by a hump. The hump remains below the baseline (Fig. 4 A, EG No 2, arrow)
or, at greater distances from the pacing site, becomes a positive spike
(Fig. 4 A, EG No 3, arrow).
Along-fiber EGs still have a biphasic
shape with an R wave preceding the intrinsic deflection.
Figures 1,2,3,4 display the computed EGs, which were simulated by applying
a local epicardial stimulus to the monoventricular model incorporating
the Purkinje network; these EGs were obtained from sites which are similarly
located, relative to the relevant isochrones, to the experimental recording
sites shown in Figures 1-4. The entire variety of morphologic features
observed in measured EGs was also found in the computed EGs and the agreement
was maintained over the entire ventricular surface. The main similarities
relate to:
a) the simple and almost invariant biphasic shape of along-fiber EGs;
b) the evolving complexity of EG waveforms recorded at increasing across-fiber
distance from the pacing site;
c) the biphasic shape in areas excited at approximately 60 ms and the
totally positive QRS complexes in the extinction regions.
Some discrepancies between experimental and simulated EGs have been
observed. For instance the QRS duration of the experimental and simulated
EGs was not expected to be the same since the model geometry used concerns
only the left ventricle and does not match the actual dog heart dimensions.
Moreover the model incorporates a simplified Purkinje network that covers
only 1/3 of the left ventricular endocardium. The septum has no Purkinje
network. This can explain the longer simulated QRS duration, compared with
the experimental ones. Other discrepancies may be due to simplifying assumptions
in the model, e.g. the flat plateau of the action potential, which does
not reproduce the initial fast repolarization typical of epicardial fibers.
In all previous simulations we used the average epicardial potential as
the reference potential. If we choose a different reference potential,
e.g. the average of the potential on the epi and endocardium or on all
the insulated boundary, we produce only a slight modification in the EG
wave forms, which does not alter the multiphasic structure of EGs (not
shown).

Figure 4. Panel A : isolated dog heart immersed in an electrolytic
tank shaped as a 10-year boy's torso, filled with a 500
cm solution. CX = circumflex coronary branch. LAD = left anterior descending
coronary branch. Ventricle was paced from star-marked site. Elliptical
contour is the 40 ms isochrone. Unipolar EGs No 1 through No 4 were recorded
from sites 1 - 4 (large dots near the epicardial elliptical isochrone at
20 msec after pacing ). Panel B: computed EGs in monoventricular model
with the epi- and endocardium in contact with conducting media. Epicardial
EGs No 1 through No 4 were computed for sites marked by dots around the
simulated epicardial isochrone at 50 msec after stimulus.
Interpretation of EG waves With a view to interpreting the multiphasic
shape of the QRS complexes in terms of intracardiac excitation events,
we simulated EGs elicited by epicardial pacing of a fully insulated
ventricle, in order to eliminate the effect of the conducting media on
potential fields and EGs. One significant feature of the fully insulated
model is that, in the split procedure that decomposes the electric
sources into an axial and a conormal component, only the axial component
of the electrical sources associated with the wave front contributes to
the extracellular potential field, while the conormal component
is silent (except for the potential jump associated with the wave front),
as predicted by the classical uniform dipole layer model . A selection
of epicardial EGs is displayed in Fig. 5; the two sets of EGs reported
in Panels A and B were obtained from 6 cross-fiber and 6 along-fiber
sites, located at increasing distance from the pacing site along two diagonal
directions.

Figure 5. Monoventricular model fully insulated. Central
epicardial stimulation. Panels A, B : superimposed, computed epicardial
electrograms from points marked by "A" and "B" dots in Panel D. Dots are
located on semidiagonal pathways along and across fiber respectively, at
increasing distance from the stimulation site. Panel C : four superimposed
epicardial electrograms related to observation points marked by dots on
a vertical line (labeled C) in Panel D ; in this region, isochrones exhibit
a dimple-like inflection. Panel D : simulated excitation time map depicting
the spread of excitation on the epicardial surface. Time interval between
successive isochrones is 5 msec.
The main features of the EGs are:
-
1.
-
a biphasic wave form for along-fiber sites. More precisely, the
EGs display a positive R wave followed by the intrinsic deflection, see
Fig. 4 Panel A .
-
2.
-
proceeding along the cross-fiber diagonal (see Fig. 5 Panel B from
left to right), we observe at first, near the pacing site, monophasic EGs
(the first and second EGs) which become progressively triphasic (the third
EG) and then pentaphasic (the fourth and fifth EGs), because the initial
portion exhibits a quite weak negative minimum (-0.1 mV) (phase 1), see
the EG enlargement in Fig. 4 Panel E, and subsequently shows 4 additional
phases: a positive hump (phase 2) followed by a negative-going segment,
a minimum (phase 3) and a spike (phase 4) before the intrinsic deflection
that terminates with a negative spike (phase 5). Finally the last (sixth)
EGs is tetraphasic because the initial weak minimum is missing.
Notice that pentaphasic EGs are found, Fig. 5 Panels B, C, in those
cross-fiber
areas where the epicardial isochrones show a dimple-like inflection i.e.
a change of curvature. At increasing distances from the pacing site, these
EGs show, before the intrinsic deflection, a spike of increasing magnitude
as the dimple-like inflection becomes more pronounced.
These results show that the main features of the measured and computed
EGs, previously described for hearts surrounded by conducting media (see
Figs. 1-4), are also present in the case of a fully insulated ventricle;
therefore, most of the polyphasic behavior of the EGs cannot be attributed
to the influence of the interfaces with the extracardiac conducting media.
The media, however, do modify the EGs to some extent, as shown later in
this Section.
The initial positive R wave, which is characteristic of along-fiber
EGs, is the electrographic counterpart of the positive areas facing those
portions of the wave front that propagate mainly along fibers (see Fig.
8 Panel A, site 2). A site reached by the wave front as it moves along
fibers exhibits increasingly positive potentials and records a positive
R wave (Fig. 8 Panel B); when the wave front crosses the site, the intrinsic
deflection is recorded. The subsequent rising phase, where the tracing
returns to the baseline, is due to the drift of the reference potential,
as previously discussed in [19,21].
The negative-going initial deflection in EGs from cross-fiber
points near the pacing site is consistent with the epicardial potential
distribution in the relevant area, which shows an increasing negativity
for an observer that moves from the resting region toward the wave front.
More details of the potential patterns in the early stages of propagation
can be found in our previous articles [20,4].
If we record EGs at increasing transverse distances from the pacing
site, we see that the initial negative-going deflection (Q wave) is progressively
replaced by a flat segment followed by a positive- going hump, whose amplitude
increases with increasing distance from the pacing site (see Fig. 6 Panel
B).
The evolving hump is in part an epicardial reflection of the rotating
deep positivity generated by those intramural portions of the wave front
which propagate along deep fibers, whose direction rotates CCW from epi-
to endocardium. In fact at all depths, two maxima are observed near those
portions of the deep wave front that propagate along the direction of deep
fibers; this produces two helical positive potential volumes that develop
CCW from epicardium to endocardium and are associated with the twisted
left-handed helical shape of the wave front (see [3,4]).
The epicardial reflection of this deep positivity produces the first, smooth
positive hump (phase 2) in cross-fiber EGs. The rule that R waves in unipolar
EGs are associated with an excitation wave approaching the observation
point (see [22]) invariably applies to locations reached
by an excitation wave front that moves mainly along fibers, but does not
apply for cross-fiber points near the pacing site . However, at
several cm from the pacing site, in some cross-fiber locations,
a spike was observed following a first, smooth hump (see Fig. 6 Panel C).
Both the hump and the spike became more and more positive with increasing
distance from the pacing site.
To ascertain whether this spike was the epicardial manifestation of
a localized axial current source, arising where the wave front is not parallel
to the local fiber direction and therefore the current sources associated
with the wave front have an axial component we analyzed the morphology
of the split form of the EGs.

Figure 6. Components of the split form of the electrograms
in the fully insulated monoventricular model. The epicardial potential
is used as the reference potential. The axial and conormal electrograms
are displayed in the left and right column respectively In first, second
and third row are reported the electrograms from sites A, B and C in Figure
6, Panel D, rspectively.
Columns in Fig. 6 show the axial and conormal EG components for the
wave trains of the full EGs displayed in Fig. 5 A, B and C. For all
cross-fiber
and along-fiber observation sites, the conormal EGs have a biphasic
shape with a positive R-wave followed by an identical downstroke, see the
right column of Fig. 6. The positive trend (R-wave) preceding the downstroke
increases in magnitude as a function of increasing excitation times, that
is, for sites at increasing distances from the stimulus location.
From the theoretical considerations about the further split of the conormal
EG developed in Methods, it follows that, for a fully insulated
ventricle, the conormal waveform is the superposition of the jump and of
the drift component of the conormal EG, see equation (22).
The drift is given by
for our reference potential; because the upstroke of v approximates
a step function, the conormal drift is nearly proportional to the area
of the excited epicardial surface at time t, therefore it is the same for
all sites, positive and monotonically increasing as a function of time.
The addition of this drift to the jump EG component results in the biphasic
structure of the conormal waveforms; in fact, from the right column of
Fig. 5, it is evident that the conormal EG consists of a moving downstroke
superimposed to a positive trend which is the same for all recording sites.
It follows from the analysis of the conormal EGs that an insulated bidomain
structure, with equal anisotropy ratio between the intra and extracellular
media, can yield only biphasic EGs, since the axial component vanishes,
and we are left with the conormal component alone. This contention is confirmed
by EG waveforms published by Simms and Geselowitz (Fig. 4 in [15]).
Another interesting consideration is that in such a structure there is
no current flowing except within the wave front itself since from equation
(22)
results proportional to
.
Therefore, in a fully insulated heart, unipolar EGs that deviate from a
biphasic waveshape indicate that the underlying bidomain has unequal anisotropic
ratio.
The axial components of the EGs (Fig. 6, left column), have widely different
shapes, depending on the geometrical interrelationships between the propagating
wave front and the local fiber direction at the various observation points.
The first row of Fig. 6 left column, relates to along- fiber sites
and shows biphasic EGs. This configuration is maintained and emphasized
when the biphasic conormal EG is added in order to obtain the full EG.
The axial waveforms for cross-fiber sites are displayed in the left
column of Fig. 6 second and third row. These EGs demonstrate that the multiphasic
shape of the full EGs originates from the axial waveforms because the full
and axial EGs share the same polymorphic configuration from mono- to pentaphasic
EGs with a positive hump followed by a spike. Adding the conormal to the
axial component to produce the full cross-fiber EGs increases the amplitude
of the intrinsic deflection, damps the negative- going trend after the
hump and enhances the monotonically increasing trend during and after the
spikes.
We recall that the axial current density
strongly depends on
since
results parallel to the unit vector
normal to the excitation wave front surface at point x; as a result
of this geometric interaction between the direction of propagation and
the local fiber direction , especially for cross-fiber sites, the
axial EGs exhibit multiphasic features.
In the more realistic conditions, in which the epi- and endocardial
surfaces are in contact with fluids, the multiphasic EGs, with humps and
spikes, are still present, although the magnitude of the epicardial EGs
is greatly attenuated relative to the insulated case.When the heart is
in contact with a conducting medium, the heart surface component
uhs(t)
arises and produces its own contingent of currents. Extension of the previous
analysis of the individual EG waveshape can be performed based on the split
form and shows that the multiple phases dispayed by the cross-fiber
EGs derive mostly from the axial EGs because the conormal EGs, a part from
an initial weakly negative-going phase, displayes a biphasic behaviour.
In the model with fluids, too, the presence of the drift associated with
the reference potential prevents us from interpreting the spike as a characteristic
marker of an underlying axial current source. To check whether the spike
is the expression of a local current source we must first remove the drift
from the EGs.
Finally the splitting analysis could be applied to the EGs wave shape
after removing the reference or drift component that reflects the
motion, during QRS, of the reference potential from near the positive to
near the negative extreme of the potential range.
Conclusions
The aim of this study was:
-
1.
-
to assess whether our monoventricular model, based on the anisotropic bidomain
representation of the heart muscle, with unequal anisotropy ratio, was
able to reproduce the variety of mono- to multiphasic unipolar electrograms
recorded from real hearts.
-
2.
-
to investigate the electrophysiology of the multiphasic unipolar EGs.
With regard to point 1) the simulated results showed that the bidomain
model, with unequal anisotropy ratio, accurately reproduces the great variety
of multiphasic shapes experimentally recorded both near the pacing site
and at distance from the pacing site.
Concerning point 2), it was shown in [21], that
some important aspects of the EGs can be explained by considering them
as the sum of two components: i) a field component that reflects
the changes of the local potential produced by an approaching or receding
wave front. ii) a reference or drift component that reflects the
motion, during QRS, of the reference potential from near the positive to
near the negative extreme of the potential range. Interpretation of the
mechanisms that generate the different wave forms, including the humps
and spikes in the multiphasic EGs, was greatly facilitated by using the
split
form of the model. This enabled us to distinguish the EG features that
reflect the perturbations of the potential field brought about by an approaching
or receding wave front from the features that reflect the variable position
of the reference potential within the potential range (drift component).
Removing the drift should be possible to re-establish consistency between
the EG wave forms and the potential distributions and suppressed artificial
spikes produced by the drift. Further work will extend the analysis to
EGs recorded from intramural and subendocardial locations, and also to
beats produced by intramural stimulation and spontaneous sinus rhythm over
the entire heart beat including also the repolarization phase.
References
[1] Colli Franzone, P. and Guerri, L., "Spreading
of excitation in 3-D models of the anisotropic cardiac tissue. I: Validation
of the eikonal model", Math. Biosci. 113:145-209, 1993.
[2] Colli Franzone, P., Pennacchio, M. and
Guerri, L., "Accurate Computation of Electrograms in the Left Ventricular
Wall",
Math. Mod. and Meth. in Appl. Sci. M3AS
10 (4): 1-32, 2000.
[3] Colli Franzone, P., Guerri, L., Pennacchio
M. and Taccardi, B., "Spread of excitation in 3-D models of the anisotropic
cardiac tissue. II: Effects of fiber architecture and ventricular geometry",
Math.
Biosci. 147: 131-171, 1998.
[4] Colli Franzone, P., Guerri, L., Pennacchio
M. and Taccardi, B., "Spread of excitation in 3-D models of the anisotropic
cardiac tissue. III: Effects of ventricular geometry and fiber structure
on the potential distribution", Math. Biosci. 151: 51-98, 1998.
[5] Geselowitz, D. B., Barr, R. C., Spach,
M. S. and Miller III, W. T., "The impact of adjacent isotropic fluids on
electrocardiograms from anisotropic cardiac muscle. A modeling study",
Circ.
Res. 51: 602-613, 1982.
[6] Geselowitz, D. B., "On the theory of the
electrocardiogram",
Proc. IEEE 77: 857-876, 1989.
[7] Geselowitz, D. B., "Description of cardiac
sources in anisotropic cardiac muscle. Application of bidomain model",
J.
Electrocardiology 25 Suppl.: 65-67, 1992.
[8] Gulrajani, R. M., "Models of the electrical
activity of the heart and computer simulation of the electrocardiogram",
CRC
Crit. Rev. Biomed. Eng. 16: 1-66, 1988.
[9] Henriquez, C. S., "Simulating the electrical
behavior of cardiac tissue using the bidomain model",
Crit. Rev. Biomed.
Engr. 21: 1-77, 1993.
[10] Henriquez, C. S., Muzikant, A. L. and
Smoak, C. K., "Anisotropy, fiber curvature, and bath loading effects on
activation in thin and thick cardiac tissue preparations: Simulations in
a three-dimensional bidomain model",
J. Cardiovasc. Electrophysiol.
7 (5): 424-444, 1996.
[11] MacLeod, R. S. and Johnson, C. R., "Map3d:
Interactive scientific visualization for bioengineering data",
IEEE
Engineering in Medicine and Biology Society 15th Annual International Conference.
IEEE Press : 30-31, 1993.
[12] Oster, H. S., Taccardi, B., Lux, R. L.,
Ershler, P. R. and Rudy, Y., "Electrocardiographic imaging: noninvasive
characterization of intramural myocardial activation from inversely-reconstructed
epicardial potentials and electrocardiograms",
Circulation: 1496-1507,
1998.
[13] Roberts, D.E. and Scher, A. M., "Effect
of tissue anisotropy on extracellular potential fields in canine myocardium
in situ",
Circ. Res. 50: 342-351, 1982.
[14] Roth, B.J., "Action potential propagation
in a thick strand of cardiac muscle",
Circ. Res. 68: 162-173, 1991.
[15] Simms, H. D. and Geselowitz, D. B.,
"Computation of heart surface potentials using the surface source model",
J.
Cardiovasc. Electrophysiol. 6: 522-531, 1995.
[16] Spach, M. S., Miller III W. T., Miller-Jones
E., Warren, R. B. and Barr, R. C., "Extracellular potentials related to
intracellular action potentials during impulse conduction in anisotropic
canine cardiac muscle",
Circ. Res. 45: 188-204, 1979.
[17] Spach, M. S., Silberberg, W. P., Boineau,
J. P. et al. "Body surface isopotential maps in normal children, ages 4
to 14 years",
Am. Heart J.72: 640-652, 1966.
[18] Streeter, D. D., "Gross morphology and
fiber geometry of the heart", In Handbook of Physiology, Vol. 1:
The
Heart, Sect. 2:
The Cardiovascular System, Berne, R. M. (Ed.)
Chapt. 4, 61-112, Baltimore, Williams and Wilkins, 1979.
[19] Taccardi, B., Lux, R. L., MacLeod, R.
S. et al. "Electrocardiographic waveforms and cardiac electric sources",
J.
Electrocardiol. 29 (Suppl.): 98-100, 1996.
[20] Taccardi, B., Macchi, E., Lux, R. L.,
Ershler, P. R. et al. "Effect of myocardial fiber direction on epicardial
potentials",
Circulation 90: 3076-3090, 1994.
[21] Taccardi, B., Veronese S., Colli Franzone,
P. and Guerri, L., "Multiple components in the unipolar electrocardiogram:
A simulation study in a three-dimensional model of ventricular myocardium",
J.
Cardiovasc. Electrophysiol. 9: 1062-1084, 1998.
[22] Wilson, F., Macleod, A. and Barker, P.,
"The distribution of the action current produced by the heart muscle and
other excitable tissues immersed in extensive conducting medium", J.
Gen. Physiol. 16: 523-556, 1933.
[23] Yamashita, Y. and Geselowitz, D. B.,
"Source-field relationships for cardiac generators on the heart surface
based on their transfer coefficients"
IEEE Trans. Biomed. Eng. 32:
964-970, 1985.