|
QT Dispersion in Unipolar Precordial Leads
Paul Kligfield, Ken W. Lee, and Peter M. Okin
Division of Cardiology, Weill Medical College of Cornell University,
and the New York-Cornell Center of New York- Presbyterian Hospital
Correspondence: Paul Kligfield, Cornell Medical
Center, 525 East 68th Street, New York, New York 10021 USA.
E-mail: pkligfi@mail.med.cornell.edu,
phone +212 746 4686, fax +212 746 8561
Abstract. Increased QT dispersion
has clinically important predictive value for arrhythmic substrate
and adverse prognosis in patients with ischemic heart disease.
However, the mechanism responsible for variability of QT intervals
on the surface ECG is unclear. Although QT dispersion on the
standard ECG is commonly related to heterogeneity of repolarization
in underlying ventricular myocardium, recent studies indicate
that QT dispersion might more simply be explained by abnormalities
of the T wave loop that result in isoelectric projection in
one or more of the precordial leads. These findings suggest
that alternative descriptive measures of morphology of the
ST segment and T wave might become useful markers of risk
in this important patient population.
Keywords: QT Interval; Repolarization; T Wave, Coronary
Artery Disease; Predictive Value
1. Introduction
Clinical interest in QT dispersion on the surface ECG
is rooted in electrophysiologic observations that regional
heterogeneity of action potentials in adjacent cardiac muscle
tissue can initiate and propagate sustained arrhythmias
[Han and Moe, 1964; Kuo et al., 1983; Antzelevitch et al.,
1991]. In the intact heart, these arrhythmias can lead to
hemodynamically unstable ventricular tachycardia or ventricular
fibrillation, each of which can cause sudden death. Because
sudden death is a major catastrophic complication of many
common cardiac disorders, including ischemic heart disease,
considerable recent attention has been focused on diagnostic
methods for the identification of patients who are at increased
risk. Regional heterogeneity of repolarization can be detected
by comparison of intramyocardial electrograms and from electrical
signals recorded directly from the surface of the heart.
Extrapolation of these findings from regional action potentials
and from epicardial electrograms to the QT intervals of
the routine ECG leads generates the working hypothesis that
QT dispersion might be an important predictor of cardiac
arrhythmias and mortality [Day et al., 1990; Pye et al.,
1994].
2. Clinical Observations of QT dispersion
Recent clinical studies have demonstrated that dispersion
of measured QT intervals from individual leads of the surface
ECG increases with disease and has important predictive
value for inducible ventricular arrhythmia, sudden death,
and other cardiac events in a wide range of cardiovascular
diseases. In chronic ischemic heart disease, prolonged QT
dispersion measured from the standard ECG has been associated
with acute ischemia [Sporton et al., 1997] and its prognostic
value has been demonstrated in patients with chronic coronary
artery disease [Zareba et al., 1994; Trusz-Gluza et al.,
1996]. In patients after myocardial infarction, prolonged
QT dispersion has been associated with the extent of myocardial
viability and with vulnerability to ventricular tachycardia
and ventricular fibrillation [van de Loo et al., 1994; Perkiömäki
et al., 1995; Glancy et al., 1996a; Perkiömäki
et al., 1997; Gabrielli et al., 1997; Oikarinen et al.,
1998].
Prolonged QT dispersion also has been found to have predictive
value in other forms of heart disease. As summarized in
a recent report, these include chronic heart failure, hypertrophic
cardiomyopathy, mitral valve prolapse, aortic stenosis,
and hypertension [Lee et al., 1998]. The link between prolonged
QT dispersion and arrhythmogenesis is highlighted by pharmacologic
studies in patients with the long QT syndrome and with torsades
de pointes [Hii et al., 1992; Priori et al., 1994]. Predictive
value of prolonged QT dispersion for inducible ventricular
arrhythmias in patients with a variety of diagnoses has
been shown to persist after adjustment for standard risk
factors, including the presence of late potentials on the
signal-averaged ECG and subnormal ejection fraction [Lee
et al., 1997].
3. Mechanisms of QT Dispersion
Despite the evident clinical usefulness, the mechanism
underlying the predictive value of observed QT dispersion
remains uncertain. QT dispersion in patients with ischemic
heart disease is commonly attributed to heterogeneity of
ventricular repolarization detected in underlying myocardium
by individual unipolar precordial leads. Because the standard
central terminal has relatively constant potential, the
unipolar precordial leads might record unique voltage changes
under individual electrodes even on the body surface. This
can be defined as "local effect". Since regional
contribution to unipolar lead voltage varies exponentially
with distance, and the heart is not situated in the center
of the thorax, it is possible that the precordial leads
preferentially record local rather than remote potential
variations. If this occurs, a plausible explanation for
the useful predictive value of measured QT dispersion is
that it reflects the heterogeneity of repolarization within
adjacent regions of myocardium that is known to be arrhythmogenic.
However, QT dispersion might also be explained simply by
interlead variation in projection of the T wave vector onto
the different precordial leads. This can be defined as a
"projection phenomenon". Once an exploring electrode
is removed from direct contact with the heart to the surface
of the body, its ability to record unique information from
subjacent tissue is limited. A unipolar electrode on the
body torso must record the underlying potential differences
from both local and remote myocardium, as represented by
the instantaneous heart vector. Low amplitude T waves and
T wave loops that are round in the horizontal plane can
produce isoelectric components of repolarization that would
result in apparent dispersion of QT measurements in the
unipolar precordial leads. It has long been recognized that
abnormal T wave loops are a feature of important coronary
artery disease [Hoffman et al., 1966], and since abnormal
repolarization is also associated with adverse prognosis
in heart disease, this could provide an alternative reason
for the observed useful predictive value of QT dispersion.
Vector projection and regional variation are not mutually
exclusive principles, and each might affect the surface
ECG in different ways. The distinction between these processes
is schematically illustrated in Fig. 1.

Figure 1. In this schematic
illustration, the terminal negative component that lengthens
the QT in V6 is isoelectrically projected onto V1. Is the
QT dispersion that results a "local effect" or a "projection
phenomenon"?
In this context, QT dispersion measured from routine ECGs
might result as much or more from isoelectric projection
of the T wave as from underlying heterogeneity of repolarization.
This would not detract from the established prognostic value
of increased QT dispersion in patients with various forms
of heart disease, but it would modify our appreciation of
the mechanism of this relationship. Accordingly, we evaluated
the potential contribution of the projection phenomenon
to the measurement of QT dispersion in normal subjects and
in patients with coronary artery disease.
4. QT Dispersion and the Projection Phenomenon
Interlead variation of ECG measurements during depolarization
and repolarization depend on the relative placement of electrodes
on the body, the distance of exploring unipolar electrodes
from the heart, and the magnitude and orientation of the
underlying electrical forces. The effect of vector projection
on interlead differences of ECG measurements can be separated
from local effects of regional heterogeneity. With respect
to QT dispersion, measurement of interlead differences of
QT interval can be made under conditions in which interlead
dispersion of repolarization is not possible. Regional heterogeneity
of QT intervals is eliminated when analysis is limited to
the bipolar and unipolar limb leads that are mathematically
derived from two bipolar frontal plane electrode pairs during
simultaneous lead acquisition of computerized ECGs [Kors
et al., 1999]. QT dispersion due to heterogeneity of repolarization
should also be greater in standard ECGs than in tracings
that are synthesized directly from three orthogonal leads
[Macfarlane et al., 1998].
To further examine the potential magnitude of apparent
QT dispersion that can be attributed to the projection phenomenon,
we eliminated local heterogeneity effects by transformation
of 12 lead ECGs into derived ECGs based on the heart vector
[Lee et al., 1998]. The analog device used to produce the
derived ECGs was developed by Dr. Gordon Dower. It first
generates an orthogonal representation of the heart vector
by transformation of the standard 12 lead ECG and then synthesizes
the derived 12 lead tracing by means of a separate algorithm.
Because the derived ECG is synthesized only from the heart
vector data, it is not merely a mathematical recreation
or restoration of the original tracing, and it contains
no information that is unique in the derived "unipolar,
precordial" leads.
We examined paired standard ECGs and derived ECGs in 129
men and women (mean age 50 years) with normal resting ECGs
that were taken prior to elective surgical procedures [Lee
et al., 1998]. We also examined paired standard and derived
ECGs in 78 hospitalized patients (mean age 67 years) with
established coronary artery disease [Kligfield et al., 1998],
most of whom had abnormal repolarization on the standard
ECG. The QT interval in each precordial lead was measured
from the first deflection of the QRS complex to the end
of the T wave, and dispersion was calculated as the difference
between maximum and minimum precordial values. When low
amplitudes of the T wave made measurement points uncertain,
individual leads were excluded from calculation to avoid
differences based on obvious measurement error.
In these studies, mean precordial QT dispersion from the
standard ECGs of patients with ischemic heart disease (51
ms) was greater than the mean QT dispersion measured from
the standard ECGs of the normal subjects (41 ms). In each
population, the average magnitudes of QT dispersion that
were measured from the derived ECGs were similar to those
measured from the standard 12 lead ECGs. Since the derived
ECGs by definition contain no heterogeneity of repolarization
duration in the precordial leads, it is evident that the
projection phenomenon can create apparent QT dispersion
in normal subjects and in patients with coronary disease
that is similar in magnitude to that measured from standard
ECGs.
5. Discussion
These findings do not exclude heterogeneity of repolarization
as a factor in the QT dispersion measured from the standard
ECG of some patients. Recognition of the projection phenomenon
in no way reduces the importance of variability of action
potential duration in arrhythmogenesis. However, these findings
do suggest that precordial QT dispersion found on the standard
ECG is not equivalent to regional heterogeneity of repolarization
in normal subjects and in some patients with coronary artery
disease.
The magnitude of QT dispersion measured in the standard
precordial leads of these patients is similar to the dispersion
that can be explained by variable precordial projection
of the T wave loop. These observations, taken together with
those of others [Macfarlane et al., 1998; Kors et al., 1999],
suggest that apparent QT dispersion may be more a measure
of T loop shape than a measure of regional heterogeneity
of repolarization. Indeed, rounding of the T wave loop in
patients with coronary artery disease and with other forms
of heart disease is evident in the vectorcardiographic work
of Hoffman et al. [1966], among others. In these studies,
a ratio of the long axis and its perpendicular transverse
dimension was used to quantify the shape of the T wave loop,
and this ratio was shown to differ in normal subjects and
in patients with ischemic disease. Abnormality of repolarization
on the ECG alone and on the vectorcardiogram is an important
sign of coronary disease and a predictor of coronary events.
The recent observation that T wave loop morphology can be
a marker for cardiac events in the elderly [Kors et al.,
1998] is consistent with these principles and observations.
It is therefore possible that the prognostic value of QT
dispersion on the surface ECG is more closely related to
its novel description of abnormal repolarization than to
its representation of local heterogeneity of action potentials.
If so, it also is possible that other quantitative approaches
to classification of ECG repolarization might provide additional
insight into the natural history of ischemic disease. Alternative,
and possibly complementary, descriptors are particularly
desirable in view of the methodologic difficulties inherent
in accurate measurement of QT intervals. Recognized problems
include accurate determination of the end of the T wave
in the presence of low amplitude T waves and precision and
reproducibility of the measurement [Kautzner et al., 1994;
Murray et al., 1994, Glancy et al., 1996b]. In this context,
principle component analysis of the T wave appears to be
a promising mathematical improvement of the early quantitative
vectorcardiographic approach noted above that is adaptable
to standard computerized electrocardiography [Xue and Reddy,
1997; Priori et al., 1997; Okin et al., 2000].
References
Antzelevitch C, Sicouri S, Litovsky SH, Lukas A, Krishnan
SC, Di Diego JM, Gintant GA, Liu DW. Heterogeneity within
the ventricular wall: electrophysiology and pharmacology
of epicardial, endocardial, and M cells. Circulation Research,
69:1427-1449, 1991
Day CP, McComb JM, Campbell R. QT dispersion:
an indication of arrhythmia risk in patients with long QT
intervals. British Heart Journal, 63:342-4, 1990.
Gabrielli F, Balzotti L, Bandiera A: QT
dispersion variability and myocardial viability in acute
myocardial infarction. International Journal of Cardiology,
61:61-7, 1997.
Glancy JM, Garratt CJ, de Bono DP. Dynamics
of QT dispersion during myocardial infarction and ischaemia.
International Journal of Cardiology, 57:55-60, 1996 (a).
Glancy JM, Weston PJ, Bhullar HK, Garratt
CJ, Woods KL, de Bono DP. Reproducibility and automatic
measurement of QT dispersion. European Heart Journal, 17:1035-9,
1996 (b).
Han J, Moe GK. Nonuniform recovery of excitability
in ventricular muscle. Circulation Research, 14:44-60, 1964
Hii JTY, Wyse DG, Gillis AM, Duff HJ, Solylo
MA, Mitchell LB. Precordial QT interval dispersion as a
marker of torsades de pointes: disparate effects of class
Ia antiarrhythmic drugs and amiodarone. Circulation, 86:1376-82,
1992.
Hoffman I, Taymor R, Kittell I. T loop rotation in ischemic
heart disease. In Vectorcardiography. Hoffman I and Taymor
RC, editors. Philadelphia and Montreal: J. B. Lippincott
Company, 1966, 181-196.
Kautzner J, Yi G, Camm AJ, Malik M. Short-
and long-term reproducibilty of QT, QTc, and QT dispersion
measurement in healthy subjects. Pacing and Clinical Electrophysiology,
17:I-928-37, 1994.
Kligfield P, Lee KW, Dower GE, Okin PM.
Is heterogeneity of repolarization required to explain QT
dispersion in patients with coronary disease? Circulation
(Suppl I) 98:I-647, 1998
Kors JA, de Bruyne MC, Hoes AW, van Herpen
G, Hofman A, van Bemmel JH, Grobbee DE. T-loop morphology
as a marker of cardiac events in the elderly. Journal of
Electrocardiology, 31 (Suppl):54-59, 1998.
Kors JA, van Herpen G, van Bemmel JH. QT
dispersion as an attribute of T-loop morphology. Circulation,
99:1458-1463, 1999.
Kuo C-S, Munakata K, Reddy P, Surawicz B.
Characteristics and possible mechanism of ventricular arrhythmia
dependent on the dispersion of action potential durations.
Circulation, 67:1356-67, 1983.
Lee KW, Kligfield P, Okin PM, Dower GE.
Determinants of precordial QT dispersion in normal subjects.
Journal of Electrocardiology, 31(Suppl):128-133, 1998.
Macfarlane PW, McLaughlin SC, Rodger JC.
Influence of lead selection and population on automated
measurement of QT dispersion. Circulation, 98:2160-2167,
1998.
Murray A, McLaughlin NB, Campbell R. Measuring
QT dispersion: man versus machine. Heart, 77:539-42, 1997.
Oikarinen L, Viitasalo M, Toivonen L. Dispersion
of the QT interval in postmyocardial infarction patients
presenting with ventricular tachycardia or with ventricular
fibrillation. American Journal of Cardiology, 81:694-7,
1998.
Okin PM, Xue Q, Reddy S, Kligfield P. Electrocardiographic
quantitation of heterogeneity of ventricular repolarization.
Journal of Noninvasive Electrocardiology, in press, 2000
Perkiömäki JS, Koistinen JM, Yli-Mäyry
S, Huikuri HV. Dispersion of QT interval in patients with
and without susceptibility to ventricular tachyarrhythmias
after previous myocardial infarction. Journal of the American
College of Cardiology, 26:174-9, 1995.
Perkiömäki JS, Huikuri HV, Koistinen
JM, Mäkikallio T, Castellanos A, Myerburg RJ. Heart
rate variability and dispersion of QT interval in patients
with vulnerability to ventricular tachycardia and ventricular
fibrillation after previous myocardial infarction. Journal
of the American College of Cardiology, 30:1331-8, 1997.
Priori SG, Napolitano C, Diehl L, Schwartz
PJ. Dispersion of the QT interval: a marker of therapeutic
efficacy in the idiopathic long QT syndrome. Circulation,
89:1681-9, 1994.
Priori SG, Mortara DW, Napolitano C, Diehl
L, Paganini V, Cantu F, Cantu G, Schwartz PJ. Evaluation
of the spatial aspects of T-wave complexity in the long-QT
syndrome. Circulation, 96:3006-3012, 1997
Pye M, Quinn AC, Cobbe SM. QT interval dispersion:
a non-invasive marker of susceptibility to arrhythmia in
patients with sustained ventricular arrhythmias? British
Heart Journal, 71:511-4, 1994.
Sporton SC, Taggart P, Sutton PM, Walker
JM, Hardman SM. Acute ischaemia: a dynamic influence on
QT dispersion. Lancet, 349:306-9, 1997.
Trusz-Gluza M, Woznaik-Skowerska I, Giec L, Szydlo K. Dispersion
of the QT interval as a predictor of cardiac death in patients
with coronary heart disease. Pacing and Clinical Electrophysiology,
19:II-1900-4, 1996.
Van de Loo A, Arendts W, Hohnloser SH. Variability
of QT dispersion measurements in the surface electrocardiogram
in patients with acute myocardial infarction and in normal
subjects. American Journal of Cardiology, 74:1113-8, 1994.
Xue Q, Reddy S. Computerized QT analysis
algorithms. Journal of Electrocardiology, 30 (suppl):181-186,
1997.
Zareba W, Moss AJ, le Cessie S. Dispersion
of ventricular repolarization and arrhythmic cardiac death
in coronary artery disease. American Journal of Cardiology,
74:550-3, 1994.
|